Ion-science SF6 P2 Manuale Utente Pagina 11

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 16
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 10
SF6 P1 & P1:p MANUAL Ion Science Ltd
10 of 15
Unrivalled Detection. www.ionscience.com
Changing the sensor unit
For changing the sensor unit, carry out the following Steps:
Power down the P2 (remove the fuse, observing the hints below)
Remove screws
Take off SmartSensor as indicated below
Pull the connector off the SmartSensor
Fit the replacement SmartSensor
Power up the P2 (insert fuse observing the hints below)
Note:
To precisely determine SF6 concentration, SF6 AREACHECK P2 requires a reference value representing
clean air. This reference value is taken and stored whenever the instrument is powered up or reset.
You should therefore make sure that power up takes place under clean air conditions.
ATTENTION: The fuse (T1A) is at mains voltage. All works on it are to be carried out by qualified
personnel only, observing appropriate precautions!
Vedere la pagina 10
1 2 ... 6 7 8 9 10 11 12 13 14 15 16

Commenti su questo manuale

Nessun commento